A fume cupboard is LEV, and COSHH treats it like any other extraction: a thorough examination and test at least every 14 months, a report kept for five years, and a label on the sash. Most schools and labs test every year so the date sits in the same holiday.

Where several cupboards are tested on one visit, expect roughly £60 to £120 per cupboard excluding VAT, with a minimum visit charge for one or two.

Per cupboard, several
£60 to £120
Test at least every
14 months
School guidance
CLEAPSS G9

Get a quote

Who needs fume cupboards tested

Secondary schools and sixth form colleges, further and higher education labs, hospital and pathology labs, pharmaceutical and chemical companies, water and environmental testing labs, and quality control labs in manufacturing.

Schools use CLEAPSS guide G9 for fume cupboards alongside HSG258. A school test is best booked in a holiday or a free period so the cupboards can be emptied and the sash run through its full range.

Recirculating (filtered) cupboards need the filter checked as well as the airflow, because a saturated filter lets fumes straight back into the room.

Typical fees for fume cupboard testing

ItemTypical fee, ex VAT
One or two cupboards
Minimum visit, report included
£150 to £300
Each cupboard, several on one visit
Ducted or recirculating
£60 to £120
Recirculating filter check
Filter test or replacement priced separately
on quote
Other lab LEV on the same visit
Extraction arms, snorkels, canopies, balance enclosures
£25 to £60

Typical UK fees read from testing firms’ published prices, November 2026. A guide, not a quote.

Estimate your fee with the cost guide

What the examiner checks on a fume cupboard

  • Face velocity

    Air speed across the sash opening, measured at a grid of points at the working sash height, and compared with the design figure.

  • Containment

    Smoke released inside the cupboard should stay inside. Draughts from doors, windows and people walking past are common causes of failure.

  • Sash and stops

    The sash should run freely and stop at the marked working height, and the alarm or airflow indicator should work.

  • Ducts and fan

    For ducted cupboards, the duct run, the fan on the roof or in the plant room, and the discharge clear of air intakes.

  • Filters

    For recirculating cupboards, the filter type for the chemicals used, and its saturation or age.

Science block cupboards due their annual test? Get a quote.

Get a quote

How a fume cupboard test runs

  1. Clear the cupboards

    the day before

    Remove stored chemicals and equipment, because a cluttered cupboard does not perform as designed and cannot be tested properly.

  2. Grid and smoke tests

    on the day

    Face velocity is measured across the sash, then smoke tests show containment.

  3. Label each cupboard

    end of the visit

    A pass label with the next due date, or a fail label and a note to stop using it.

  4. Report to the science team

    after the visit

    Keep the report with the logbook, and give the head of science or lab manager a copy.

Asked before booking

How often should a school fume cupboard be tested?

At least every 14 months under COSHH. Schools usually test every year, often in the summer holiday.

Do recirculating fume cupboards need testing?

Yes, and the filter needs checking too. The test covers airflow and containment, and the report should say whether the filter is right for the chemicals used.

Can the cupboard be used while waiting for a repair?

Not one with a fail label. Move the work to a cupboard that passed, or postpone it.